UMF · Model 4010 Power Procedure Table
4010 Power Procedure Table One of our most versatile and fully featured power procedure tables, the UMF Medical is ADA compliant, lowers to a height of 19 inches (48.3cm) and can support a patient weighing up to 600 pounds (272.2kg). The Ultra Comfort Seamless Top provides security and comfort for patients and also allows for easy cleaning and helps prevent cross-contamination. The 4010 Power Procedure Table with its four function hand control, is an excellent choice for family general practices as well as in OB/GYN, dermatology, plastic surgery, urology and proctology. The low access height, versatility, and superior weight capacity of the 4010 Power Procedure Table make it a perfect table for a variety of exams, procedures, and minor surgeries. The UMF Medical 4010 Power Procedure Table features a fully articulating power assisted back section, headrest, and leg section.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Height, tilt, back or leg section will not move from the pendant / foot control, or drifts under load.
Root cause: Linear actuator, a hand / foot control fault, a control-board fault, or (hydraulic) a valve / seal leak.
Symptoms: Intermittent or dead control, one function unresponsive.
Root cause: Worn control or a damaged cable.
Symptoms: Float / longitudinal / lateral drive will not move or lock, radiolucent top damaged.
Root cause: Drive motor, brake, or the float-release solenoid.
Symptoms: No movement on a power interruption.
Root cause: Aged back-up battery.
Symptoms: Torn pads, loose stirrups / arm boards, paper-roll holder broken.
Root cause: Worn upholstery or accessory hardware.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Linear actuator | Major service part | $$$ | Difficult |
| Hand / foot control | Periodic replacement | $$ | Fairly easy |
| Control board | Major service part | $$$ | Moderate |
| Hydraulic pump / seal kit (hydraulic tables) | Major service part | $$ | Moderate |
| Upholstery / pad set | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio