Tuttnauer · Model 3850EP
General Safety Features. Double Independent Monitoring: The combined electronic and mechanical monitoring guarantees accurate results. The operator has two independent means to monitor temperature and pressure. Safety Valves: The chamber is equipped with safety valves – if the pressure exceeds the allowed limit the safety valves will discharge. Built-in Steam Generator Safety: A water level monitoring system maintains a constant water level and ensures safe operation of the heaters Control Features. High precision control system for perfect sterilization results. Password protection allows for secure access control. Independent temperature and pressure monitoring. Cycle information recovery in the case of power failure or cycle interruption. Fail Alert – Indicates cycle failure or interruption. Door Alert – Indicates the door is unlocked. RS 232 PC Connection Port for direct software updates and remote maintenance. Automatic, Digital Control; Chamber Diameter x Depth: 15 x 20-Inch; Capacity: 65-Liters; Includes support stand and 2 Stainless Steel Tray Shelves (14 x 20-Inch). Includes: 60-Minute Timer, Integrated Printer, 3.25-Liter Reservoir & Low Water-Level Alarm. Temperature Range: 105 to 137°C (221 to 279°F)Digital display of temperature & pressure. Heating time to 273°F (134 °C): 30-MinutesHeat-insulated, double-locking door protects against accidental burns. Door cannot be opened while chamber is under pressure. Automatic low water level shut-off prevents damage to heating elements when there is insufficient water for the sterilization cycle. UL Listed, CSA Certified & ASME Stamped.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Cycle aborts, temperature or pressure not reached / held, failed BI or chemical indicator.
Root cause: Heater / element fault, a door-gasket leak, a steam-generation or water-fill fault, a blocked chamber drain / strainer, or a temperature / pressure-sensor fault.
Symptoms: Steam leak at the door, door will not seal, lock or unlock, pressure will not build.
Root cause: Worn door gasket or a door-lock / interlock fault.
Symptoms: Instruments / packs come out wet.
Root cause: Drying-phase vacuum / heater fault, an overloaded chamber, or a clogged drain.
Symptoms: Low-water alarm, mineral scale, cloudy chamber, poor steam quality.
Root cause: Empty or scaled reservoir, a fill-valve / pump fault, or use of non-distilled water.
Symptoms: Failed Bowie-Dick / air-removal test, slow come-up, air pockets.
Root cause: Blocked steam trap, a vacuum-pump / ejector fault, or a check-valve fault.
Symptoms: No or faint cycle printout.
Root cause: Thermal print head or paper.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Door gasket | Periodic replacement | $$ | Fairly easy |
| Heating element / steam-generator element | Major service part | $$$ | Difficult |
| Chamber / feedwater strainer & fill valve | Periodic replacement | $$ | Moderate |
| Temperature / pressure sensor | Major service part | $$ | Moderate |
| Steam trap / solenoid & check valves | Periodic replacement | $$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio