Thermo Fisher Scientific · Model MaxQ 8000
Three times the capacity of a floor shaker in virtually the same footprint Built for both sensitive eukaryotic culture applications - with contamination-reducing HEPA filtration, coved corner design, single-piece chamber and tight temperature uniformity - and for demanding prokaryotic culture applications Rugged construction, wide speed range and no limits on “hard shake” speed, even when stacked three-high.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Platform will not shake, speed does not match set point, drive fault.
Root cause: Door / lid not closed, an eccentric-drive or belt fault, an out-of-balance load, or a controller fault.
Symptoms: Chamber below (or above) set point, temperature alarm.
Root cause: Heater element, a temperature sensor, a control-board fault, or (refrigerated) a compressor / sealed-system fault.
Symptoms: Displayed temperature disagrees with an independent probe.
Root cause: A programmed offset, a defective RTD, or a main-board fault.
Symptoms: Rumble or knocking while shaking.
Root cause: Load imbalance, or a worn eccentric bearing / drive component.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Eccentric-drive bearing / belt | Major service part | $$ | Difficult |
| Heater element | Major service part | $$ | Difficult |
| RTD temperature sensor | Major service part | $$ | Moderate |
| Drive motor / controller | Major service part | $$$ | Difficult |
| Refrigeration compressor (refrigerated models) | Major service part | $$$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio