EPPENDORF · Model New Brunswick Innova 42/42R
Incubator shaker The Innova 42 incubated and 42R incubated and refrigerated shakers are our most versatile incubator shakers. The programmable Innova 42 and 42R provide exceptional convenience allowing you to simultaneously shake and static incubate cultures under the same conditions. Use on or under the bench, or double stacked for space savings. An optional culture drawer, gassing manifold, remote alarm, and humidity monitor are available for both models; a photosynthetic light bank and UV germicidal light are available for the 42R.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Platform will not shake, "Open Door" icon on display, no response to start.
Root cause: Power cord unplugged / switch off, door not closed firmly, an improperly seated fuse after a replacement, on/off switch fault, or shaking speed set to zero by a running program or the computer interface.
Symptoms: Chamber below (or above) set point, temperature alarm.
Root cause: Blown heater fuse or compressor fuse (42R), compressor over-pressure switch activated, ambient too hot/cold, a defective heater, or a defective refrigeration system (42R).
Symptoms: Displayed temperature does not match an independent probe.
Root cause: A programmed Temperature Offset (Offset icon shown), a defective RTD assembly, or a defective main board.
Symptoms: Rumble or knocking during shaking.
Root cause: Load out of balance on the platform, or a loose component in the platform / subplatform / drive assembly.
Symptoms: Chamber will not cool below ambient, compressor short-cycles or trips.
Root cause: Compressor, over-pressure switch, or sealed-system fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Reinstall carefully - a poorly seated fuse mimics a dead unit.
Sealed-system work; refrigeration certification required.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater fuse / compressor fuse Reinstall carefully - a poorly seated fuse mimics a dead unit. | Major service part | $ | Fairly easy |
| RTD temperature sensor assembly | Major service part | $$ | Moderate |
| Heater element | Major service part | $$ | Difficult |
| Drive belt / eccentric bearing | Major service part | $$ | Difficult |
| Refrigeration compressor (42R) Sealed-system work; refrigeration certification required. | Major service part | $$$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.