Thermo Electron LED GmbH · Model Megafuge ST4F Plus-MD
Megafuge ST4F Plus-MD 208-240V/50/60Hz.
Intended use: An electrically-powered device intended to be used to separate the components of multiple types and various volumes of clinical suspensions using centrifugal force. It is a self-contained unit with an electric motor, a vertical shaft, and a horizontal rotor attached to the upper end. It is typically used to centrifuge various types of clinical specimens, either alone or after addition of reagents or other additives, for subsequent in vitro diagnostic analysis. It may additionally have refrigeration capabilities to perform temperature-sensitive procedures.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error message when starting, run will not begin.
Root cause: Set speed exceeds the rotor's admissible speed, an imbalanced load, or an unapproved rotor found by rotor detection.
Symptoms: Slow spin-up, rumble or vibration.
Root cause: Load imbalance, worn anti-vibration mounts, or a drive/bearing fault.
Symptoms: Electric lid release does not work (especially after a power failure).
Root cause: Lid latch/solenoid fault or a power failure - use the mechanical emergency release only after the rotor has fully stopped.
Symptoms: Chamber does not reach set temperature.
Root cause: Compressor, condenser or door-seal fault.
Symptoms: Visible pitting or a crack on a rotor or bucket.
Root cause: Age, chemical exposure or fatigue - a safety-critical item.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Lid latch / solenoid | Major service part | $$ | Moderate |
| Rotor / buckets | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.