Accuray Incorporated · Model 1057727
Software, Version 1.0.0.4, Treatment Delivery Console, Radixact (Operating Software for Radixact Treatment Delivery System)
Intended use: An assembly of devices designed to produce high-energy electrons to yield high-energy x-rays used for therapeutic applications, primarily to treat cancer. It provides a uniform radiation field of uniform intensity and predictable energy level, in a beam with well defined dimensions. The principle of operation is linear acceleration of electrons by means of electromagnetic (EM) microwaves. The output is used as a treatment beam on the patient, either directly or after passing a suitable target. The system typically includes signal analysis and display devices, patient and device supports, and software.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Beam-on faults, dose-rate low or unstable, RF/HV interlocks.
Root cause: Magnetron at end of life, HV supply, or waveguide arcing.
Symptoms: Leaf stuck open or closed, delivery interlock, QA fluence errors.
Root cause: Pneumatic leaf actuator or valve fault in the binary multileaf collimator.
Symptoms: Rotation error, data or power dropouts during a helical delivery.
Root cause: Slip-ring brushes or the ring drive.
Symptoms: Poor daily image quality for IGRT setup, registration difficulty.
Root cause: Imaging source or detector aging or calibration drift.
Symptoms: TDC will not connect to the delivery system, plan will not load, session cannot start.
Root cause: Console PC, network, or software fault.
Symptoms: Over-temperature interlocks, dose-rate instability.
Root cause: Chiller pump or heat-exchanger fouling.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
OEM service; dosimetry re-commissioning required.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Magnetron OEM service; dosimetry re-commissioning required. | Periodic replacement | $$$$ | Expert-level |
| Binary MLC valve bank / leaves | Major service part | $$$$ | Expert-level |
| kVCT / MVCT detector | Major service part | $$$$ | Difficult |
| Slip-ring brushes | Periodic replacement | $$ | Difficult |
| Chiller pump / heat exchanger | Major service part | $$$ | Difficult |
| Treatment-delivery console PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.
"MLC tickle error" may result in the delivered dose to effectively rotate from the planned dose.
There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
Uncontrolled couch Z-axis movement (descent)