Terumo · Model Terufusion TE-311
Ease of use: Your best guarantee for a faultless and SAFE infusion [b]Display for safe parameter settings[/b] In order to avoid reading mistakes and misprogramming of the flow rate, thedifference between the digits and the decimal places is clearly marked by an orange coloured decimal point and an apparent difference in size. [b]Unique dial setting[/b] The ergonomically designed dial helps to eliminate mistakes whilst making settings. It allows you to reach the setting easily and in a more evident manner.Fast and unmistakable syringe placementIn order to guarantee correct placement and reliable functioning of the syringe, the pump is equipped with a barrel detection unit. In this way, the correct placement of different syringe sizes is very fast and easy, as well as entirely safe. [b]Solid design[/b] Especially during patient transportation, it is important to have a system which is designed to minimize the risk of breaking. Therefore, it is a big advantage to have the Terufusion syringe pumps where the slider is always staying inside the borders of the pump casing. [b]Clear alarm monitor[/b] This monitor centralizes information about the operation status and nature of the alarm. The status can easily be seen on the pump, without having to enter menus. A pressure monitor is incorporated in the alarm monitor to follow the pressure variance constantly. A large, bright Operation Indicator features a unique flow-proportional indication for intuitive monitoring of the flow. With its omnidirectional view, it enhances safety in terms of flow rate setting (green turning), Stand-by Mode (green flashing) and alarm status (red).
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.