Terumo · Model Sarns 7400
Heart-Lung Bypass Unit Complete Open Heart System with Pulsatile
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Poor gas exchange or temperature control.
Root cause: Oxygenator membrane fouling or heat-exchanger leak.
Symptoms: Inaccurate flow, unusual noise from the pump head.
Root cause: Raceway wear or centrifugal rotor bearing wear.
Symptoms: False or missed air-in-line/level alarm.
Root cause: Ultrasonic sensor drift or dirty optics.
Symptoms: Console freeze or loss of readouts mid-case.
Root cause: Main control board or display module fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Sterilizable disposable raceway; console head inspected at PM.
Single-use per case.
Safety-critical; verify at PM.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Console main board | Major service part | $$$$ | Expert-level |
| Pump head / raceway Sterilizable disposable raceway; console head inspected at PM. | Major service part | $$$$ | Difficult |
| Oxygenator (disposable) Single-use per case. | Consumable | $$$$ | Fairly easy |
| Level / bubble sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.