Stockert · Model CAPS
Computer aided perfusion system (CAPS) Mobile, stainless steel console base including two telescoping masts. Compact, streamline, easy-to-reach controls. Can be used with or without a computer.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Poor gas exchange or temperature control.
Root cause: Oxygenator membrane fouling or heat-exchanger leak.
Symptoms: Inaccurate flow, unusual noise from the pump head.
Root cause: Raceway wear or centrifugal rotor bearing wear.
Symptoms: False or missed air-in-line/level alarm.
Root cause: Ultrasonic sensor drift or dirty optics.
Symptoms: Console freeze or loss of readouts mid-case.
Root cause: Main control board or display module fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Single-use per case.
Safety-critical; verify at PM.
Sterilizable disposable raceway; console head inspected at PM.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Oxygenator (disposable) Single-use per case. | Consumable | $$$$ | Fairly easy |
| Level / bubble sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
| Console main board | Major service part | $$$$ | Expert-level |
| Pump head / raceway Sterilizable disposable raceway; console head inspected at PM. | Major service part | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →