Smiths Medical · Model CADD Prizm PCA 6000 Series
• Easy to read display • Reliable and Durable • Programming flexibility • Air-in-Line detector • Event memory log • Lock levels for patient safety
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Uncontrolled flow when door open.
Root cause: Worn roller or failed free-flow valve.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable; per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
| IV administration set Consumable; per patient. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
In certain cases, the lock and key supplied with Graseby 3300 pumps and accessories does not match the standard Graseby 3300 key.
One lot of CADD High Volume Administration sets, part number 21-7381-24, lot number 26X035, incorrectly contained the IFU insert for part number 21-7357-24.
Product temperature readings may be 2C higher than the actual temperature
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.