• Easy to Read Display - Easy to read and understand display screen provides information for programming adjustments and troubleshooting • Cassette Detection - Cassette detection sensor is designed to detect cassette attachment • Occlusion Detection - Upstream and downstream occlusion sensors are designed to alert the patient and clinician if interruption of fluid delivery occurs • Air-in-Line Detection - Air-in-line detector with Off, High- and Low-sensitivity settings is designed to alert the patient and clinician if air is present in the administration set tubing • ON/OFF Key - ON/OFF key allows pump to be placed in low-power state and is designed to save battery life • Event Memory - The last 1,000 events in the pump's history can be downloaded either directly or remotely (using a modem) to a PC using the CADD-DIPLOMAT® PC Communications System • Reliability - CADD® pump reliability and durability can mean less down time and fewer interruptions to therapy
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Certain CADD Legacy 1, Ambulatory Infusion Pumps, may contain non-matching serial numbers. The label attached to the back of the pump may list a serial number that does not match the correct serial number programmed into the software of the pump.
The CADD-Legacy 1 Model 6400 Ambulatory Infusion Pump (Serial Number 370969) was shipped without documentation that the required accuracy testing had been performed during the manufacturing process. If the subject infusion pump did not go through the accuracy testing, and if it does not meet accuracy specifications, there is a potential for over or under-delivery.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →