Sinothinker · Model SK201
Windows operation interface, figure button, mouse operation 96-well plate, multiple tests in the single plate. Arrangement of Test item in micro-plate can be store and re-use Parameter can be modified and stored Plate shaking, speed and time adjustable Open system, test item and parameters can be added or modified. Large storage for 1000 test item and 1,000,000 results Printing comprehensive report, include hospital and patient information and test results.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blank or standard absorbance off, poor plate-to-plate reproducibility, high CV.
Root cause: Lamp aging, a filter-wheel fault, or contamination on the optics.
Symptoms: Plate not loading, jams, wells read at the wrong position.
Root cause: Worn transport belt / motor, a position-sensor fault, or an out-of-spec plate.
Symptoms: Shake step does not run or is uneven.
Root cause: Shaker drive fault.
Symptoms: Software will not connect, or an unresponsive display.
Root cause: USB / serial link or the display board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Lamp | Periodic replacement | $$ | Moderate |
| Filter wheel / interference filters | Major service part | $$ | Moderate |
| Plate transport belt and motor | Periodic replacement | $$ | Difficult |
| Plate position sensor | Periodic replacement | $ | Moderate |
| Shaker drive | Major service part | $$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.