Siemens Healthineers · Model Acuson SC2000
Auto Left Ventricular Package for LV quantification consists of automated reference plane extraction from the cardiac volume dataset and automated volume contouring of ejection fraction. Auto Reference Plane Extraction automatically generates and displays images corresponding to standard 2D views or planes from a volume dataset in the Left Ventricular Analysis (LVA) application and Rapid Stress volume stress echo application improving efficiency and consistency. Auto Volume Contouring (within the LVA application) automatically traces the endocardial borders of multiple reference planes and calculates standard measurements eliminating manual tracing. Streamlined natural and extended reach zones puts frequently used controls at user fingertips and enables quick access to peripheral devices, recordable media and transducer connectors. The control panel is designed for motor-memory learning. Configurable desk format with 90 degree side-to-side positioning enhances user comfort and proximity to patient. Flexible articulating arm and high-resolution 20.1-inch wide screen flat-panel display for optimal monitor placement and enhanced image quality in all lighting situations. Ergonomic 4Z1c volume imaging transducer features a patented Palmar grip with an elastomeric gripping surface to facilitate a positive hand grip while scanning. To further enhance user efficiency, the syngo SC2000 Workplace workstation enables complete review and processing of the acquired volumes - freeing the ultrasound system for acquisition.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.