Infant / Pediatric Ventilator From Sechrist Industries comes the Millennium® - the infant ventilator and pediatric ventilation system designed for maximum performance and user-friendly convenience. The Millennium® Infant Ventilator combines state-of-the-art electronics with the simplicity of pressure sensing technology. The result: simple, reliable ventilation for neonatal, infant, and pediatric patient requirements. Reliable, Versatile, Simple. Neonatal Ventilation modes such as SIMV/IMV, Assist Control, and Backup ventilation in CPAP make the Millennium® Neonatal Ventilator the optimal choice in basic, reliable infant ventilation and versatile pediatric ventilation. What sets the Millennium® Infant Ventilator apart is the patented SmartSync™, a unique pressure sensing device that synchronizes the patient's natural breathing cycle with the neonatal ventilator. It has a maximum sensitivity of 0.10 cmH20, and a response time of 5-40 milliseconds depending upon the sensitivity settings. With this technology from a company you've known and trusted for decades, you can be certain the Millennium® Infant Ventilator and Pediatric/Neonatal Ventilation system will provide your patients with unparalleled performance.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Self-test fault, circuit-leak or compliance-check failure, will not be released for use.
Root cause: Patient-circuit / expiratory-valve leak, a flow-sensor fault, an O2-cell fault, or an internal pneumatic fault.
Symptoms: Delivered tidal volume or flow does not match the setting, or fails the annual verification.
Root cause: Flow-sensor drift / contamination, an inspiratory / expiratory valve fault, or a blender / proportioning-valve fault.
Symptoms: Low-supply alarm, FiO2 not matching the setting, O2-cell reads wrong.
Root cause: Wall / cylinder supply issue, an inlet filter / regulator fault, the blender, or an aged O2 cell.
Symptoms: PEEP drifts, auto-PEEP, high-pressure alarms, poor exhalation.
Root cause: Worn or wet expiratory valve / diaphragm, or a blocked exhalation path.
Symptoms: Short internal-battery run time, will not charge, transfers to battery unexpectedly.
Root cause: Aged internal battery or a power-supply fault.
Symptoms: Noisy operation, cannot meet flow demand, turbine / compressor alarm.
Root cause: Worn turbine / compressor.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor(s) | Frequent wear item | $$ | Fairly easy |
| Expiratory valve / diaphragm | Frequent wear item | $$ | Fairly easy |
| O2 cell / sensor | Frequent wear item | $$ | Easy |
| Internal battery | Periodic replacement | $$ | Fairly easy |
| Inlet filters / water traps | Frequent wear item | $ | Easy |
| Turbine / compressor (turbine units) | Major service part | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.