• Easy on-screen monitoring in real-time or remotely using EasyCare Tx software • Remote alarm II and connectivity to central monitoring hospital system with dedicated cable • In-built Learn Circuit which at the push of a button evaluates and compensates for any resistance in the circuit • The ResMed Leak Valve includes an anti-asphyxia valve, allowing safe use in adult or paediatric patients, with both cuffed or un-cuffed tracheotomies • Disease-specific, pre-set Pathology Defaults enable quicker and easier therapy initiation - allowing your care provider to fine-tune pressure settings required for you specifically
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Low exhaled tidal volume, high leak alarm, inability to maintain PEEP.
Root cause: Diaphragm/valve degraded by secretions and cleaning cycles.
Symptoms: Inaccurate delivered volume, false volume/pressure alarms.
Root cause: Sensor contamination or aging of differential-pressure transducer.
Symptoms: Unable to reach set pressure, low-flow alarm, loud noise.
Root cause: Worn bearings or failed blower motor.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Bacterial / HEPA filter Consumable - replace per PM schedule. | Consumable | $ | Easy |
| Compressor / turbine module | Periodic replacement | $$$ | Difficult |
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
| O2 sensor (galvanic/paramagnetic) | Frequent wear item | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.