Reichert Technologies · Model UNISTAT Bilirubinometer
Measures total bilirubin levels in under 5 seconds. Most trusted and reliable stat analysis of serum bilirubin available. Measures undiluted serum, eliminating errors from time-consuming dilution or chemical reactions. Easy to calibrate using included optical standards. Correlates between 96-99% with HPLC gold standard for measuring bilirubin.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Result disagrees with a reference / control, drift over time.
Root cause: Optics contamination, lamp aging, a dirty or scratched cuvette, or calibration drift.
Symptoms: Low signal, will not read, lamp warning.
Root cause: Lamp at end of life.
Symptoms: Bubble or short-sample errors, inconsistent readings.
Root cause: Air bubble in the micro-sample, or a worn / dirty cuvette.
Symptoms: Unresponsive keypad, dim display.
Root cause: Aged display or keypad.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Lamp | Periodic replacement | $$ | Moderate |
| Cuvette | Frequent wear item | $ | Easy |
| Optical filter / detector assembly | Major service part | $$$ | Difficult |
| Calibration standard | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.