Dräger · Model JM-103
Reusable Probe - Just wipe with alcohol and reuse as needed. There are no disposable tips to fumble with, keep track of or reorder. Portable Convenience - Compact device fits easily in hand or pocket and allows for quick clean-up between uses. Ergonomic Design - Controls assure convenient operation while helping seal out contaminants and extend device life. Easy-to-read Display - See results in seconds, not hours. Large numbers help ensure accurate interpretation. Results are shown in mg/dL or µmol/L as selected by user.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "Measured value is abnormal" or large fluctuation during average measurement.
Root cause: Probe not held perpendicular to the recommended site (forehead or sternum), poor skin contact, or a measurement error during averaging.
Symptoms: Warning persists after power-cycle (OFF, wait 10 s, ON).
Root cause: Internal hardware fault - requires a Konica Minolta Sensing authorized service facility.
Symptoms: Low-battery indication, short run time, will not power on.
Root cause: Meter not seated on the charger unit, or an aged rechargeable battery.
Symptoms: Readings drift, inconsistent results, visible marks on the probe tip.
Root cause: Dirty or scratched reusable probe optical window (wiped with alcohol between uses).
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable battery | Periodic replacement | $$ | Fairly easy |
| Charger / cradle unit | Major service part | $$ | Easy |
| Measuring probe / optical head | Major service part | $$$ | Moderate |
| Check / calibration standard | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio