APEL · Model BR-5200
Very quick measurement - Max. one second measurement time. Computer connection - Has an RS-232C and a USB interface for connection with computer. No lamp replacement - Does not need a lamp replacement thanks to the LED light source. Free of calibration hassle - Once BR-5200 is calibrated, no more daily calibration is required. Easy set of the sample tube - A sample gauge provided on the surface of BR-5200, which indicates the proper sample volume to be inspected. Only simple action is required to set the capillary tube into the holder. Handles hemolysis and turbidity easily - The dual wavelength measuring method of BR-5200 eliminates interference of hemolysis and turbidity.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Result disagrees with a reference / control, drift over time.
Root cause: Optics contamination, lamp aging, a dirty or scratched cuvette, or calibration drift.
Symptoms: Low signal, will not read, lamp warning.
Root cause: Lamp at end of life.
Symptoms: Bubble or short-sample errors, inconsistent readings.
Root cause: Air bubble in the micro-sample, or a worn / dirty cuvette.
Symptoms: Unresponsive keypad, dim display.
Root cause: Aged display or keypad.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Lamp | Periodic replacement | $$ | Moderate |
| Cuvette | Frequent wear item | $ | Easy |
| Optical filter / detector assembly | Major service part | $$$ | Difficult |
| Calibration standard | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio