RARaypa · Model ID Culture
Incubation by natural convection. Setting temperature range from room temperature +5°C to 80°C. Heating by electric elements assembled in an independent chamber for optimal temperature stability. Temperature and time controlled by digital PID microprocessor, with 10 editable programs of customizable name, programmable with up to 10 temperature/time segments. Timer start and timer stop adjustable by date, hour and minute. Intuitive alphanumeric 3,5’’ LCD touchscreen user interface that displays time, temperature and cycles in graphs, has audiovisual alarms and advanced cycle and timer programming.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Chamber above or below setpoint, temperature alarm, slow recovery after the door opens.
Root cause: Heater element, a temperature-sensor / controller fault, a door / gasket leak, or a circulation-fan fault.
Symptoms: CO2 above or below setpoint, CO2 alarm, pH drift in cultures.
Root cause: CO2 sensor (TC or IR) drift, an empty / low CO2 supply, or a solenoid / regulator fault.
Symptoms: Low humidity, plates drying out, condensation problems.
Root cause: Empty water pan, or an evaporation / airflow issue.
Symptoms: Chamber stops heating, over-temp alarm, will not restart.
Root cause: Controller / sensor fault driving the heater high, or an independent over-temp thermostat.
Symptoms: Door will not seal, temperature swings, condensation at the door.
Root cause: Worn door gasket or a sagging hinge.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater element | Major service part | $$ | Moderate |
| Temperature sensor / controller | Major service part | $$ | Moderate |
| CO2 sensor (CO2 incubator) | Periodic replacement | $$$ | Moderate |
| CO2 solenoid / regulator (CO2 incubator) | Major service part | $$ | Moderate |
| Door gasket | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.