Puritan Bennett · Model Achieva
Offers both volume and pressure control ventilation. Flow and/or Pressure triggering for enhanced patient synchrony. Internal dial-in PEEP takes the guesswork out of setting pressure trigger sensitivity. Pressure Support to enhance weaning and improve patient comfort and patient ventilator/synchrony. Expiratory Sensitivity improves patient synchrony by allowing the clinician to adjust the percent of peak inspiratory flow that will cycle the ventilator from inspiration to expiration in Pressure Support. Exceptional battery life for extended mobility. Internal O2 blending for accurate delivery of 21% to 100% oxygen. (PSO2 model). Flow Acceleration Limit. Internal memory and modem allow access to stored data, providing a tool to troubleshoot alarms and evaluate effectiveness of patient settings. Can be used with pediatric patients as small as 5 kg.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Self-test aborts with a diagnostic code, circuit-leak or compliance test fails, ventilator not released for use.
Root cause: Patient-circuit leak, an exhalation-valve / flow-sensor fault, or an internal pneumatic / electronic fault (see the diagnostic-code table).
Symptoms: FiO2 not matching the setting, low or high supply-pressure alarm, gas-mix alarm.
Root cause: Blender fault, a wall-supply / hose issue, or a failing internal compressor / turbine.
Symptoms: Monitored tidal volume, pressure or flow implausible, frequent nuisance alarms.
Root cause: Flow sensor(s) or the internal pressure transducers aging or contaminated.
Symptoms: PEEP not held, auto-PEEP, high circuit pressure, exhalation-valve alarm.
Root cause: Worn or wet exhalation-valve diaphragm / assembly.
Symptoms: Short battery run time, will not ride through a power interruption, will not charge.
Root cause: Aged internal / external battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor(s) | Frequent wear item | $$ | Fairly easy |
| Exhalation valve assembly | Frequent wear item | $$ | Fairly easy |
| O2 sensor (galvanic cell) | Frequent wear item | $$ | Fairly easy |
| Internal compressor / turbine | Major service part | $$$$ | Expert-level |
| Battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.