Puritan Bennett · Model 560
• The internal lithium ion battery provides up to 11 hours of ventilator operation (depending on settings and other factors).* • A real-time battery life indicator shows how much battery time is left, in hours and minutes, based on settings. • The Puritan Bennett™ 560 ventilator is compact, lightweight (4.5 kg) and quiet (<30 dBA at 1 m). It transitions easily from hospital to home use. • Data portability is enabled via a USB memory device. The USB memory device makes it easy to transfer patient data from a home ventilator to a clinicians computer. • Ventilator parameters and alarm settings are linked to reduce the risk of clinically inappropriate settings. • Oxygen enrichment is available using a low pressure oxygen source. Utilize a measurement kit to monitor FiO2 levels
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Low exhaled tidal volume, high leak alarm, inability to maintain PEEP.
Root cause: Diaphragm/valve degraded by secretions and cleaning cycles.
Symptoms: Inaccurate delivered volume, false volume/pressure alarms.
Root cause: Sensor contamination or aging of differential-pressure transducer.
Symptoms: Unable to reach set pressure, low-flow alarm, loud noise.
Root cause: Worn bearings or failed blower motor.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable, replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Bacterial / HEPA filter Consumable, replace per PM schedule. | Consumable | $ | Easy |
| Compressor / turbine module | Periodic replacement | $$$ | Difficult |
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
| O2 sensor (galvanic/paramagnetic) | Frequent wear item | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Directions for Use-Puritan Bennett has determined that a service interval should be defined for the device. The firm has also determined that an exhalation filter should be employed when the device is in use.
Manufacturing defect at contract manufacturer, employee error, poor training, improper inspection.