Critical care ventilator The volume controlled and pressure controlled ventilation with mandatory and assist breaths supports patient needs with care. With user-friendly keypad, wide 9” screen and scroll-n-select knob, it takes just minutes for anyone to start ventilating a patient with Inventa. All the critical parameters concerned to the mode of ventilation are available at a glance, which helps to make effective and faster clinical decisions.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Inaccurate delivered volume, false volume/pressure alarms.
Root cause: Sensor contamination or aging of differential-pressure transducer.
Symptoms: Unable to reach set pressure, low-flow alarm, loud noise.
Root cause: Worn bearings or failed blower motor.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: Leak alarm, low circuit pressure, volume loss.
Root cause: Disconnected or punctured patient circuit / humidifier.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Most-replaced wearable part; stock spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
| O2 sensor (galvanic/paramagnetic) | Frequent wear item | $$ | Fairly easy |
| Exhalation valve assembly Most-replaced wearable part; stock spares. | Frequent wear item | $$ | Fairly easy |
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.