Philips Healthcare · Model Volcano Core Mobile
True Integration and Convenience Flexibility to service multiple rooms Plug-and-play simplicity of digital IVUS Touchscreen control from the sterile field to get your answer is faster Hyperemia-free iFR modality One System, Many Choices FFR lesion assessment Optimal ease of use Guided workflows Uniform controls to simplify training Measurements and labeling tools.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, dim or absent intravascular image, or an unstable / drifting FFR pressure.
Root cause: Imaging catheter / pressure wire damage, a patient-interface-module (PIM) fault, or an acquisition-engine fault.
Symptoms: Automated pullback does not run at the set rate, stalls, or the distance is wrong.
Root cause: Pullback motor / sled or the PIM coupling.
Symptoms: System does not recognise the device, calibration / normalisation fails.
Root cause: PIM connector wear or a damaged catheter / wire hub.
Symptoms: No image to the boom monitor, angio co-registration misaligns, export fails.
Root cause: Video routing, network, or workstation fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient interface module (PIM) / pullback sled | Periodic replacement | $$$ | Fairly easy |
| Imaging catheter / pressure wire | Consumable | $$$ | Easy |
| Acquisition engine board | Major service part | $$$$ | Difficult |
| Workstation PC / display | Major service part | $$$ | Moderate |
| Cath-lab video / co-registration interface | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.