
Philips Healthcare · Model M3015A CO2 MMS Module
Microstream CO2 is a high performance Sidestream CO2 measurement with a very low sample flow rate that can be used on patients ranging from neonates to adults. A wide variety of Microstream accessories is available for intubated and non-intubated patients, for humidified or non-humidified ventilation, and for CO2 measurement with or without supplemental Oxygen (O2) delivery. Attached to the MMS, it provides a lightweight and rugged package that can be mounted to the patient monitor, or positioned close to the patient to reduce cable clutter and can be transferred from one monitor to another to ensure seamless data transfer with the patient. All patient connections are fully electrically isolated individually for maximum patient safety.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: CO2, O2, agent or tcpO2/tcpCO2 values off vs a reference, calibration prompts.
Root cause: Sensor aging, a missed calibration, or the wrong calibration gas; (tcpO2) an old membrane / electrolyte.
Symptoms: Host monitor (IntelliVue / MP series) does not display the module or MMS.
Root cause: Docking / rack connector or cable fault.
Symptoms: Occlusion or water-trap alarm, no or low sample flow, flat waveforms.
Root cause: Full or expired water trap, or a blocked / kinked sample line.
Symptoms: Intermittent signal with cable movement.
Root cause: Worn patient cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set | Frequent wear item | $$ | Easy |
| Water trap / sample line | Consumable | $ | Easy |
| O2 / CO2 sensor or gas-bench service | Frequent wear item | $$ | Fairly easy |
| Module / MMS docking connector | Major service part | $$ | Moderate |
| Internal sample pump | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio