Spacelabs · Model 91517 CO2 Module
CO2 Module The 91517 Capnography module uses both main and side stream gas to analyze and measure the carbon dioxide in the gas mixture. This determine ventilatory, circulatory, and metabolic status. The Co2 sensor is self calibrating which prevents interruption of connection while monitoring. Units uses airway adapters, endotracheal, and or tracheostomy tubes while using the mainstream function.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Host monitor does not recognise the module.
Root cause: Interface PCBA fault, missing supply rails (+5 V at J1-2, +12 V at J1-11, -12 V for SDLC), or a flash/software error on the interface board.
Symptoms: Message will not clear after 7 minutes.
Root cause: Failed CO2 sensor - swap with a known-good sensor.
Symptoms: Message with no waveform.
Root cause: Both CO2 sensors disconnected or a cable fault.
Symptoms: After a brief power loss the module resets in ~20 s and returns to default settings.
Root cause: Known issue - the module resets 20 s after power failure instead of 3 minutes, clearing RAM (subject of a recall).
Symptoms: EtCO2 offset or slow response.
Root cause: Fouled or wet airway adapter, or the mainstream sensor window contaminated.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Mainstream CO2 sensor | Periodic replacement | $$$ | Fairly easy |
| Interface PCBA | Major service part | $$$ | Difficult |
| Airway adapters | Consumable | $ | Easy |
| Sidestream sampling line / water trap | Consumable | $ | Easy |
| Module housing / latch | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings following power on or reset.
Capnography Module, Model 91517, resets 20 seconds after power failure, instead of after three minutes as specified. This reset clears the RAM and the module returns to default settings.
There have been reports that the Spacelabs Medical Vital Signs Monitor, Model 91220, will not power up with AC power and will no longer charge the batteries.
The transducer used to compensate for pressure changes was mounted backwards on the printed circuit board assembly. The module will not compensate for pressure changes and CO2 readings will be inaccurate at higher altitudes.