CO2 Module The 91517 Capnography module uses both main and side stream gas to analyze and measure the carbon dioxide in the gas mixture. This determine ventilatory, circulatory, and metabolic status. The Co2 sensor is self calibrating which prevents interruption of connection while monitoring. Units uses airway adapters, endotracheal, and or tracheostomy tubes while using the mainstream function.
Synced live from the FDA openFDA API.
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings following power on or reset.
Capnography Module, Model 91517, resets 20 seconds after power failure, instead of after three minutes as specified. This reset clears the RAM and the module returns to default settings.
There have been reports that the Spacelabs Medical Vital Signs Monitor, Model 91220, will not power up with AC power and will no longer charge the batteries.
The transducer used to compensate for pressure changes was mounted backwards on the printed circuit board assembly. The module will not compensate for pressure changes and CO2 readings will be inaccurate at higher altitudes.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.