Philips Healthcare · Model M1027A
Modules The EEG module M1027A produces two channels of EEG realtime waves EEG1 and EEG2 which can be displayed and recorded. The M1027A EEG Module monitors the patient’s cerebral function by measuring the electrical activity of the brain.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, NIBP, temperature or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a module input-board fault.
Symptoms: Host monitor (IntelliVue / MP series) does not display the module or MMS.
Root cause: Docking / rack connector or cable fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: Intermittent signal with cable movement.
Root cause: Worn patient cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Module input board | Major service part | $$$ | Difficult |
| Module / MMS docking connector | Major service part | $$ | Moderate |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.