Philips Healthcare · Model M1018A TCPO2/TCPCO2 Module
Module Philips Transcutaneous Gas is designed for use in neonatal intensive care environments. Skin surface tcpO2/tcpCO2 measurements correlate with changes in arterial partial pressure of oxygen and carbon dioxide, offering a non-invasive means of continuous monitoring. Individual and trend data provided by transcutaneous gas monitoring are a significant complement to arterial blood gas measurements in monitoring neonates on ventilators.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: CO2, O2, agent or tcpO2/tcpCO2 values off vs a reference, calibration prompts.
Root cause: Sensor aging, a missed calibration, or the wrong calibration gas; (tcpO2) an old membrane / electrolyte.
Symptoms: Host monitor (IntelliVue / MP series) does not display the module or MMS.
Root cause: Docking / rack connector or cable fault.
Symptoms: Occlusion or water-trap alarm, no or low sample flow, flat waveforms.
Root cause: Full or expired water trap, or a blocked / kinked sample line.
Symptoms: Intermittent signal with cable movement.
Root cause: Worn patient cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set | Frequent wear item | $$ | Easy |
| Water trap / sample line | Consumable | $ | Easy |
| O2 / CO2 sensor or gas-bench service | Frequent wear item | $$ | Fairly easy |
| Module / MMS docking connector | Major service part | $$ | Moderate |
| Internal sample pump | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.