Regulate anesthetic drug use. Decreases the incidence of post-operative side effects such as nausea and vomiting. Reduce length of stay in the PACU (recovery room). Prevents intraoperative awareness. Greater awareness of patient-specific reaction to anesthesia and sedation. BISx power link works with a range of sensors.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: High SQI / EMG, noisy or flat EEG, BIS / index value unreliable or absent.
Root cause: Sensor contact / expiry, the patient interface cable (PIC), or a DSC / module fault.
Symptoms: Host monitor does not display the module.
Root cause: Docking / rack connector or cable fault.
Symptoms: Intermittent signal with cable movement.
Root cause: Worn patient / sensor cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| BIS sensor / patient interface cable | Frequent wear item | $$ | Easy |
| Digital signal converter (DSC) | Major service part | $$$ | Fairly easy |
| Module / rack docking connector | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.