Philips Healthcare · Model Azurion 3 F15
Enhance Visibility for Vessels - the intricacies of complex malformations, tumors and occlusions, as well as less radio-opaque devices are fully visible Save Time - team members can work on all activities in the control and exam room without interrupting each other Take Control of Dose - has a comprehensive suite of radiation dose management tools, training and integrated product technologies so you can take control of patient care, staff safety, and regulatory compliance Insightful Live Image Guidance - helps clinicians determine the most beneficial course of treatment with confidence Evolve as Your Needs Change - you can easily integrate extra functionality and third-party applications.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or artefacted fluoro / DSA images, poor detail.
Root cause: X-ray tube aging, flat-panel detector degradation, or a collimator / filter fault.
Symptoms: Gantry rotation, angulation or the stand will not move or gives a position error / collision stop.
Root cause: Drive motor, encoder, brake, or a collision-sensor fault.
Symptoms: Table float, tilt or height will not move or drifts.
Root cause: Table drive / brake, the tabletop float solenoid, or a control fault.
Symptoms: Exposure inhibited, kV / mA error, tube-overload or thermal interlock.
Root cause: Generator fault, tube thermal limit, or the cooling system.
Symptoms: Review workstation or the in-room monitor blank, no images to PACS.
Root cause: Video routing, a workstation fault, or a network / PACS-interface fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| X-ray tube assembly | Major service part | $$$$ | Expert-level |
| Flat-panel detector | Major service part | $$$$ | Expert-level |
| C-arm / stand drive motor & encoder | Major service part | $$$$ | Expert-level |
| Table drive / float solenoid | Major service part | $$$ | Difficult |
| Tube cooling unit / heat exchanger | Major service part | $$$ | Difficult |
| Review workstation / video-routing hardware | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio