
Pentax · Model EB-1970UK
High-resolution ultrasound and remarkable endoscopic imaging due to a fully digital video and 100% color CCD chip in the tip technology. The convex ultrasound probe allows sonographic visualization of the bronchial system for physicians in the diagnosis and staging of lung cancer, especially lymph nodes and tumor masses. Fast orientation, navigation, and precise positioning of the EBUS endoscope is possible due to the US probe always being visible on screen. Parallel monitoring of endoscopic and ultrasound image through picture-in-picture mode. Color and Power Doppler for optimized characterization. Large 2.0mm diameter of the instrument channel provides good suction capability and facilitates the easy insertion of devices.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: The manual or automated leak test fails; bubbles from the bending section, insertion tube, or a channel port.
Root cause: A pinhole in the bending-rubber, a channel breach, or a damaged distal cap - fluid invasion risk.
Symptoms: Blurred, dark, spotty, half-image or no image; colour cast that will not white-balance.
Root cause: Fluid invasion into the CCD / fibre bundle, a broken image bundle, or lens-surface damage / adhesive fog.
Symptoms: Up/down or left/right angulation is loose, will not hold, over-tight, or limited range.
Root cause: Stretched or broken angulation wires, a worn brake, or bending-section damage.
Symptoms: Poor suction, no or weak air / water, will not pass a cleaning brush.
Root cause: Debris or a kink in the channel, a worn air/water or suction valve seat, or a crushed channel.
Symptoms: Dim or no illumination from the scope with a known-good light source.
Root cause: Broken light-guide fibres or a damaged light-guide connector.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Bending-rubber / distal cap | Frequent wear item | $$ | Moderate |
| Angulation wires / brake assembly | Periodic replacement | $$$ | Difficult |
| Air/water & suction valves / seats | Frequent wear item | $ | Easy |
| Image bundle / CCD repair (OEM/3rd-party) | Periodic replacement | $$$$ | Expert-level |
| Light-guide bundle / connector | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio