OES Medical · Model Cygnus
Anaesthesia Ventilator Combines ease of use with sophisticated modes Large clear colour touch screen Volume / Pressure, Spontaneous, PSV – Plus, SIMV and SMMV modes Dual Waveforms High quality easy to use with low running costs Intuitive touch screen control with large easy to read colour screen Integrated Oxygen monitor with pop up key pad adjustment Variable orifice based Spirometry Electronic PEEP 2 hour battery backup Patient select initial screen User selectable patient weight default settings for 5, 10, 25, 50, 100 and 150 Kg User saveable patient configured settings with custom naming facility – six off Patient setting confirm button Touch screen set buttons for ventilator settings Pop up key pads for adjusting ventilator settings – Oxygen alarm, Minute volume, Tidal volume, Pressure limit and Target, PEEP, Rate – BPM, and I: E Ratio Measured value display box Colour coded wave forms Mode button display box with second row prompt arrow Ventilation mode select button – usable in case with user confirm Mute button with 120 second delay and reset facility Gas mix option with pop out menu Language options Time Case time with stop and reset facility.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Self-test fault, circuit-leak or compliance-check failure, will not be released for use.
Root cause: Patient-circuit / expiratory-valve leak, a flow-sensor fault, an O2-cell fault, or an internal pneumatic fault.
Symptoms: Delivered tidal volume or flow does not match the setting, or fails the annual verification.
Root cause: Flow-sensor drift / contamination, an inspiratory / expiratory valve fault, or a blender / proportioning-valve fault.
Symptoms: Low-supply alarm, FiO2 not matching the setting, O2-cell reads wrong.
Root cause: Wall / cylinder supply issue, an inlet filter / regulator fault, the blender, or an aged O2 cell.
Symptoms: PEEP drifts, auto-PEEP, high-pressure alarms, poor exhalation.
Root cause: Worn or wet expiratory valve / diaphragm, or a blocked exhalation path.
Symptoms: Short internal-battery run time, will not charge, transfers to battery unexpectedly.
Root cause: Aged internal battery or a power-supply fault.
Symptoms: Noisy operation, cannot meet flow demand, turbine / compressor alarm.
Root cause: Worn turbine / compressor.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor(s) | Frequent wear item | $$ | Fairly easy |
| Expiratory valve / diaphragm | Frequent wear item | $$ | Fairly easy |
| O2 cell / sensor | Frequent wear item | $$ | Easy |
| Internal battery | Periodic replacement | $$ | Fairly easy |
| Inlet filters / water traps | Frequent wear item | $ | Easy |
| Turbine / compressor (turbine units) | Major service part | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.