Datex Ohmeda · Model Ohmeda 7000
Calibrated electronic control allows for a more accurate ventilation delivery. Minute volume, I: E ratio and rate controls can adjust without affecting other setting Comprehensive alarms and safety features. Accommodates nearly any patient size.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Self-test fault, circuit-leak or compliance-check failure, will not be released for use.
Root cause: Patient-circuit / expiratory-valve leak, a flow-sensor fault, an O2-cell fault, or an internal pneumatic fault.
Symptoms: Delivered tidal volume or flow does not match the setting, or fails the annual verification.
Root cause: Flow-sensor drift / contamination, an inspiratory / expiratory valve fault, or a blender / proportioning-valve fault.
Symptoms: Low-supply alarm, FiO2 not matching the setting, O2-cell reads wrong.
Root cause: Wall / cylinder supply issue, an inlet filter / regulator fault, the blender, or an aged O2 cell.
Symptoms: PEEP drifts, auto-PEEP, high-pressure alarms, poor exhalation.
Root cause: Worn or wet expiratory valve / diaphragm, or a blocked exhalation path.
Symptoms: Short internal-battery run time, will not charge, transfers to battery unexpectedly.
Root cause: Aged internal battery or a power-supply fault.
Symptoms: Noisy operation, cannot meet flow demand, turbine / compressor alarm.
Root cause: Worn turbine / compressor.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor(s) | Frequent wear item | $$ | Fairly easy |
| Expiratory valve / diaphragm | Frequent wear item | $$ | Fairly easy |
| O2 cell / sensor | Frequent wear item | $$ | Easy |
| Internal battery | Periodic replacement | $$ | Fairly easy |
| Inlet filters / water traps | Frequent wear item | $ | Easy |
| Turbine / compressor (turbine units) | Major service part | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The electronic vaporization system in the GE Aisys Anesthesia machine contains the following components: Backpressure Valve, Inflow Check Valve and Cassette Interface Board, that can contribute to independent failures.
Neonatal Flow Sensors supplied for use with the Neonatal Enhancement for the Engstrm Carestation may result in the sensors providing invalid values and alarm messages when exposed to high flow rates.
Certain Compact Absorbers may have an increased resistance to gas flow due to an improperly manufactured foam filter. The increased resistance can cause an elevated pressure at the ventilator end of the inspiratory circuit, but the pressure at the patient may be reduced. This could result in patient hypoventilation and hypoxia.
Vaporizer shutdown with alarm-An intermittent leak in the Aladin 2 cassette/Aisys pneumatic interface (valves) may occur that will invoke an alarm state. When this occurs 'Vaporizer Failure' will appear on the Aisys main display, with an accompanying audible alrm. Delivery of anesthetic Agent is halted; gas flow and ventilation are not affected.
Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.
Phototherapy medical device may lose its position and can cause a safety hazard to the patient.
Anesthetic agents could be delivered at concentrations above the concentration set on the S/5 and AS/3 Anesthesia Delivery Units (ADU) having serial numbers within a specified range.
Datex-Ohmeda Cardiocap/5 Patient Monitors have bracket pins which do not adequately secure them to their optional mounting stand so the monitors may fall.
Possible disruption of mechanical ventilation or disruption of anesthetic agent delivery.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio