OES Medical · Model Carina
Inspiratory time 0.25 – 3.0 seconds Expiratory time 0.5 – 6.0 seconds Flow Rate 0.1 – 1.0 litres per second Adjustable Pressure Relief 20 – 80 cm H2O I:E Ratio derived from inspiratory and expiratory time minimum 1:1 ratio Tidal Volume 25ml – 1600ml Adult Bellows 25ml – 300ml Paediatric Bellows Driving Gas Oxygen Supply Pressure 45 – 100 PSI (310 – 700 KPA) Airway Pressure Guage -20 – 100cm H2O Range Dimensions H x W x D (mm) 340 x 300 x 260 mm Weight 8 kg Adjustable Low Pressure Alarm 0 – 30cm H2O Adjustable High Pressure Alarm 10 – 70cm H2O Mute Button with 15 second delay and LED
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: O2 low-pressure alarm, inability to ventilate.
Root cause: Pipeline disconnect or failed O2 flush valve.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: Rising EtCO2, patient rebreathing.
Root cause: CO2 absorbent saturated.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace on schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Breathing circuit / bellows | Periodic replacement | $$ | Fairly easy |
| CO2 absorber canister Consumable - replace on schedule. | Consumable | $ | Easy |
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →