ndd · Model EasyOne Pro
Single Breath CO diffusion (DLCO). Spirometry. Automatic calibration. Flexible EMR capabilities. No warm-up time. Challenge testing. No expensive service contracts. High-resolution color touch screen.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: 3-L syringe verification outside tolerance, questionable FEV1 / FVC.
Root cause: Debris or condensation in the ultrasonic flow path, a damaged FlowTube / spirette, or transducer aging.
Symptoms: DLCO, TLC or FRC results drift, gas calibration fails.
Root cause: Gas analyser aging, a calibration-gas issue, a sample-line leak, or the demand valve / gas delivery.
Symptoms: Software will not connect, no data transfer.
Root cause: USB cable / port or driver.
Symptoms: Leak at the mouthpiece, contamination flags, poor manoeuvres.
Root cause: Worn valve, or a reused single-use FlowTube / filter.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
The ultrasonic sensor has no moving parts.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| FlowTube / spirette (single-use) The ultrasonic sensor has no moving parts. | Consumable | $ | Easy |
| Calibration gas | Frequent wear item | $$ | Easy |
| Communication cable | Periodic replacement | $ | Easy |
| Breathing valve / demand valve (Pro) | Periodic replacement | $$ | Fairly easy |
| 3-L calibration syringe | Major service part | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio