FVC (Forced Vital Capacity), SVC (Slow Vital Capacity), and MVV (Maximum Voluntary Maximum) test Pre- and post-bronchodilator comparison Accurate measurement by using disposable mouthpiece with built-in sensor PFT test report with interpretation.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Calibration verification fails against a 3-L syringe, FVC/FEV1 drift.
Root cause: Debris or moisture in the flow path, a damaged disposable mouthpiece sensor, or use of non-authorized accessories causing noise.
Symptoms: Unit will not power on.
Root cause: Blown 250 V 5 A high-breaking fuse at the AC power connection port, a power-cord fault, or a power-switch fault.
Symptoms: No printed PFT report, faint print, paper jam, printer cover switch not detected.
Root cause: Worn thermal print head, empty/jammed paper, or a printer-cover switch fault.
Symptoms: Intermittent signal, ECG or spirometry input drops.
Root cause: Worn connector at the patient-cable connection port or the spirometry port.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-patient consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Mains fuse (250 V, 5 A, high-breaking) | Major service part | $ | Fairly easy |
| Disposable mouthpiece with built-in sensor Single-patient consumable. | Consumable | $ | Easy |
| Thermal printer print head and paper | Periodic replacement | $ | Fairly easy |
| Patient cable | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.