High definition colour touch-screen. Simple screen icon interface using stylus or optional MicroMouse. Fast on-screen textual entry enabling comments and report writing. Fast, quiet printer with easy load paper mechanism. Fully customisable printout format. Direct connection to external Hewlett Packard printers. Click here to view a list of compatible printers. Unique contextual Help Screens (language selectable) with dedicated Help Button. 41 inspiratory and expiratory indices measurable. On-screen graphical child incentive device. Large 2000 patient test memory. Internationally recognised on-screen test quality assurance prompts (e.g. slow start, abrupt end etc.) ensuring quality test results. Unique todays testing feature - creating a list of patients tested or to be tested. Numerous predicted values and languages selectable. Lung Age assessment and textual result interpretation by ATS (9), BTS-NICE (10) or Enright (11). Body Mass Index and Dyspnoea Score. Bronchodilator and/or Steroid assessment. Complete with all accessories in a sturdy carrying case.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: 3-litre syringe calibration verification outside tolerance, questionable FEV1/FVC.
Root cause: Physically damaged or contaminated volume transducer, or debris / moisture in the flow path.
Symptoms: Dead unit or frequent low-battery.
Root cause: Depleted battery or a battery-contact fault.
Symptoms: Drifting results, visible soiling in the flow head.
Root cause: Reusable transducer needs cleaning, or a disposable mouthpiece reused.
Symptoms: No printout, faint print, paper jam.
Root cause: Worn print head or paper handling.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Calibration is stable unless the transducer is physically damaged.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable volume transducer Calibration is stable unless the transducer is physically damaged. | Periodic replacement | $$ | Easy |
| Disposable mouthpieces / bacterial filters | Consumable | $ | Easy |
| Battery | Frequent wear item | $ | Easy |
| Printer paper / print head | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.