Streamlined Injection Workflow - Less time preparing for the injection and more time to focus on the patient and optimize procedure management. Seamless Network Communication with Radimetrics Enterprise Platform for Contrast Dose Management Connect with the Radimetrics™ Enterprise Platform for automated documentation and viewable patient histories to enhance operational consistency. Convenience at Point of Care - On-board eGFR and Weight Based Dosing Calculators, an Injection Pressure Graph and independent Test Inject and KVO functions are convenient features located right on the Injection System. Deliver with Confidence in Quality - The next generation in MR consumables Bayer stands behind. Designed, tested and warranted as part of the Total MR SMART Injection System. Improved Efficiencies For Workflow and Automated Fluid Delivery Snap-on/twist-off syringe design Automatic plunger advance and retract when attaching and detaching syringes Automatic filling and priming Injection pressure graph Personalized Care Through Convenience at Point of Care Patient-Centric workflow design Modality worklist connectivity Protocol storage/retrieval. On-board eGFR and Weight Based Dosing Calculators Injection enabled when head is tilted down Reproducible Quality To Enhance Operational Consistency Proven company track record of design and performance Automated data capture Interface with PACS/RIS/SR Export DICOM secondary capture to PACS.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error message or code on the control panel, injector will not arm or aborts an injection.
Root cause: An operator-induced condition (syringe/air check) or a subsystem fault - the message/code indexes the troubleshooting section.
Symptoms: Piston will not move, inaccurate flow or volume, stalls under pressure.
Root cause: Plunger drive motor, drive belt, or the plunger-position / mechanical-stop potentiometer.
Symptoms: No response, blank display, or arms but no delivery.
Root cause: Power supply, control board, or a head-to-console cable fault.
Symptoms: Injector disarms near the magnet or will not run out of the down position.
Root cause: Tilt sensor or the MR-environment safety interlock.
Symptoms: Control panel text corrupt or the remote shows nothing.
Root cause: Display card or the head/console communication path.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Plunger drive motor | Major service part | $$$ | Difficult |
| Plunger motor drive belt | Periodic replacement | $$ | Moderate |
| Plunger-position / mechanical-stop potentiometer | Major service part | $$ | Moderate |
| Tilt sensor / MR safety interlock | Major service part | $$ | Moderate |
| Disposable syringe set | Consumable | $ | Easy |
| Display / controller card | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.