MedRad · Model Mark V ProVis
Height adjustable arm for ease in loading syringes Rotating, tilting control panel is easy to see and read from almost any position in the room Multiple syringe configurations (60/150 ml - 150/150 ml - 200/200 ml) help to match the right syringe size with the patient and procedure Floor, ceiling or wall mount options add flexibility Segmented control panel is easy to learn and operate "One Touch" enable button and variable speed control for fast loading and easy delivery Exclusive mechanical stop to minimize risk of over-injection bullet Articulating arm extends reach for ease in positioning close to patient.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error message or code on the control panel, injector will not arm or aborts an injection.
Root cause: An operator-induced condition (syringe/air check) or a subsystem fault - the message/code indexes the troubleshooting section.
Symptoms: Piston will not move, inaccurate flow or volume, stalls under pressure.
Root cause: Plunger drive motor, drive belt, or the plunger-position / mechanical-stop potentiometer.
Symptoms: No response, blank display, or arms but no delivery.
Root cause: Power supply, control board, or a head-to-console cable fault.
Symptoms: Contrast not warmed, heater indicator off or flashing.
Root cause: Failed syringe heat maintainer or its cable.
Symptoms: Control panel text corrupt or the remote shows nothing.
Root cause: Display card or the head/console communication path.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Plunger drive motor | Major service part | $$$ | Difficult |
| Plunger motor drive belt | Periodic replacement | $$ | Moderate |
| Plunger-position / mechanical-stop potentiometer | Major service part | $$ | Moderate |
| Syringe heat maintainer | Periodic replacement | $$ | Fairly easy |
| Disposable syringe set | Consumable | $ | Easy |
| Display / controller card | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.