MedRad · Model Mark V Plus
Pedestal, used, angiography injector. Also available in rack mount model with choice of mounting options-overhead counterpoise system, head pedestal or table mount. Complete with remote, head, control, start switch, syringe warmer, power cable and interconnect cables.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error message or code on the control panel, injector will not arm or aborts an injection.
Root cause: An operator-induced condition (syringe/air check) or a subsystem fault - the message/code indexes the troubleshooting section.
Symptoms: Piston will not move, inaccurate flow or volume, stalls under pressure.
Root cause: Plunger drive motor, drive belt, or the plunger-position / mechanical-stop potentiometer.
Symptoms: No response, blank display, or arms but no delivery.
Root cause: Power supply, control board, or a head-to-console cable fault.
Symptoms: Contrast not warmed, heater indicator off or flashing.
Root cause: Failed syringe heat maintainer or its cable.
Symptoms: Control panel text corrupt or the remote shows nothing.
Root cause: Display card or the head/console communication path.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Plunger drive motor | Major service part | $$$ | Difficult |
| Plunger motor drive belt | Periodic replacement | $$ | Moderate |
| Plunger-position / mechanical-stop potentiometer | Major service part | $$ | Moderate |
| Syringe heat maintainer | Periodic replacement | $$ | Fairly easy |
| Disposable syringe set | Consumable | $ | Easy |
| Display / controller card | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.