Medical International Research · Model Spirolab II
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: 3-L syringe calibration check fails, FVC/FEV1 drift, erratic loops.
Root cause: Debris or moisture in the turbine flow head, a damaged turbine, or use of a non-MIR turbine.
Symptoms: Nothing on power-up on internal battery.
Root cause: Internal battery pack below ~6 V, or an internal power-supply fault (5 V rail).
Symptoms: Error on the display or printer during the ~3 s power-on self-test.
Root cause: A subsystem flagged during self-test - refer to the functional-test section.
Symptoms: No printout, faint print, paper jam.
Root cause: Worn print head or paper handling.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Use only MIR turbines; disposable turbines are single-patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable turbine flow head (MIR) Use only MIR turbines; disposable turbines are single-patient. | Periodic replacement | $$ | Easy |
| Internal battery pack (6-8 V) | Periodic replacement | $$ | Moderate |
| Battery charger / power supply | Periodic replacement | $ | Fairly easy |
| Thermal print head and paper | Periodic replacement | $ | Fairly easy |
| 3-L calibration syringe | Major service part | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.