Matachana · Model STEAM STERILIZERS S2000
GENERAL FEATURES: Technology: new control system more powerful and modern. TFT touch screen for loading area with integrated PLC, as well as register PLC which ensure independent readings between temperature and pressure. Frontal made of an innovative material that prevents dirt and reflections. Easy to clean. Frontal backlit indicator icons that show the status of the sterilizer. Ergonomics: more accessible displays with a modern and versatile interface. Sustainability: energy and water consumption reduced thanks to incorporated economizer systems. Efficiency: redesigning of vacuum system by Venturi ensuring a low water consumption. Compatibility: perfect drying tested with the most demanding packaging on the market. Connectivity: ready to connect with the most usual management and traceability systems via Ethernet port. Comfort: new range of more functional accessories, adapted to the loading requirements. Accessibility: full circuits and components redistribution that facilitates the accessibility of the Technical Service for easy maintenance procedures. Remote diagnosis: remote connection via Ethernet port to know the sterilizer’s condition and enable the Technical Support Service to diagnose incidences/failures remotely for faster solution. Suitable for hospital sterilization centres, pharmaceutical laboratories and the healthcare products industry, as well as any other centre that requires fast and safe sterilization of large volume.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Cycle aborts, temperature or pressure not reached / held, failed BI or chemical indicator.
Root cause: Heater / element fault, a door-gasket leak, a steam-generation or water-fill fault, a blocked chamber drain / strainer, or a temperature / pressure-sensor fault.
Symptoms: Steam leak at the door, door will not seal, lock or unlock, pressure will not build.
Root cause: Worn door gasket or a door-lock / interlock fault.
Symptoms: Instruments / packs come out wet.
Root cause: Drying-phase vacuum / heater fault, an overloaded chamber, or a clogged drain.
Symptoms: Low-water alarm, mineral scale, cloudy chamber, poor steam quality.
Root cause: Empty or scaled reservoir, a fill-valve / pump fault, or use of non-distilled water.
Symptoms: Failed Bowie-Dick / air-removal test, slow come-up, air pockets.
Root cause: Blocked steam trap, a vacuum-pump / ejector fault, or a check-valve fault.
Symptoms: No or faint cycle printout.
Root cause: Thermal print head or paper.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Door gasket | Periodic replacement | $$ | Fairly easy |
| Heating element / steam-generator element | Major service part | $$$ | Difficult |
| Chamber / feedwater strainer & fill valve | Periodic replacement | $$ | Moderate |
| Temperature / pressure sensor | Major service part | $$ | Moderate |
| Steam trap / solenoid & check valves | Periodic replacement | $$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.