Maquet Critical Care AB · Model 6481779
Compressor Mini 115V 60Hz.
Intended use: A mains electricity (AC-powered) device designed to create compressed air and supply it directly to gas-powered or compressed air dependent devices (e.g., ventilators or other respiratory devices, laboratory equipment). It may also provide medical air depending upon the filtering systems employed.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Ventilator air-supply alarm, compressor runs but output pressure low, slow recovery.
Root cause: Worn piston / valve, an air leak in the tank or distribution, or a saturated air-dryer / filter.
Symptoms: Water in the ventilator inlet filter, condensation.
Root cause: Air-dryer cartridge saturated or an auto-drain fault.
Symptoms: No output, compressor hot, thermal overload trips.
Root cause: Failed motor / run capacitor, or a cooling-fan fault.
Symptoms: Compressor cycles at the wrong pressures, delivered pressure high or low.
Root cause: Worn pressure switch or a drifting regulator.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Piston / valve rebuild kit | Periodic replacement | $$ | Difficult |
| Air-dryer / inlet filter | Frequent wear item | $ | Fairly easy |
| Motor run capacitor | Major service part | $ | Moderate |
| Cooling fan | Periodic replacement | $ | Fairly easy |
| Pressure switch | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio