iCRco · Model iCR7200
The iCR7200 Dual Bay CR System defines a whole new class in the arena of multi loader CR solutions. 2 separate CR reading modules allow for simultaneous scanning of 2 plates resulting in an ultra high throughput of up to 140 14x17 plates per hour. Both modules are based on the patented iCRco design incorporating the True Flat Scan Path™, an ultra precise scan mechanism and protective cassettes to produce over 300,000 high resolution images per plate. iCRco CR Imaging Plates are fixed to a rigid back panel. Nothing ever touches the active area of the phosphor plate resulting in artifact free images throughout the lifetime of the CR Plate. Unlike most CR products, the iCR7200 Dual Bay CR Unit uses no rollers to transport the plate across the 2 scan heads. A positive traction drive system ensures perfect artifact free reproduction with every scan. The iCR7200 Dual Bay Unit has been designed to be deployed in Hospitals, Orthopedic facilities and in any other environment where ultra high throughput is needed. Enjoy the advantages of a multi loader CR system at the price of a single plate CR.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Banding, streaks, ghost images, non-uniform density, dust artefacts.
Root cause: Dirty light guide / scanning optics, a worn erase-lamp array, imaging-plate wear, or contaminated transport rollers.
Symptoms: Plate jams in the reader, transport error, plate not returned.
Root cause: Worn or contaminated transport rollers / belts, or a sensor fault.
Symptoms: A faint previous image appears on the next exposure.
Root cause: Erase-lamp array degradation or a driver fault.
Symptoms: No or very noisy image, scan-error message.
Root cause: Scanning laser or the photomultiplier / detection chain.
Symptoms: Images will not display, annotate or send to PACS.
Root cause: Workstation PC or the DICOM configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transport rollers / belts | Periodic replacement | $$ | Moderate |
| Erase-lamp array | Periodic replacement | $$$ | Moderate |
| Imaging plates / cassettes | Periodic replacement | $$$ | Easy |
| Light guide / scanning optics | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio