Hologic · Model Aquilex
A simple user interface that is easy to setup. Setup in three easy steps once the system is connected and powered on. One integrated system with a fluid pump capable of 800 ml/min inflow. Two integrated vacuum pumps (1 for the underbuttock drape and scope outflow, 1 for the MyoSure system) capable of producing up to 500 mmHg of suction. Automatic lumen calibration to optimize intrauterine pressure for achieving and maintaining distention and visualization with a wide variety of hysteroscopes.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure droops or overshoots, poor distension, pressure or flow alarm.
Root cause: Pump-head / roller wear, a pressure-sensor fault, a leak in the tubing / inflow-outflow, or a blocked cannula.
Symptoms: Cassette / tubing not recognised, will not prime, leaks at the pump head.
Root cause: Cassette-detection switch or worn pump-head seals.
Symptoms: Poor outflow, cannot clear debris, outflow alarm.
Root cause: Outflow pump / valve or a blocked suction line.
Symptoms: Blade will not oscillate / rotate, weak cut, handpiece not recognised, footswitch dead.
Root cause: Worn shaver handpiece / motor, or its cable / footswitch.
Symptoms: Unresponsive touchscreen or keys, will not power on.
Root cause: Display / control board or power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump-head rollers / seals | Periodic replacement | $$ | Moderate |
| Pressure sensor / transducer | Major service part | $$ | Moderate |
| Cassette-detection switch | Periodic replacement | $$ | Moderate |
| Shaver handpiece / motor (shaver systems) | Periodic replacement | $$$ | Moderate |
| Footswitch | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio