GE HealthCare · Model Marquette 200SL Multiparameter Module
Marquette 200SL Multiparameter Module The GE Tram multi-parameter module allows patient data to move with the patient throughout the hospital. This allows critical data to remain with the patient so you can access complete information at the bedside when you check on patients after surgery.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, temperature, NIBP or CO intermittently drops or is implausible.
Root cause: Worn patient cable / connector, or a module input-board fault.
Symptoms: Host monitor does not display the module or Tram / pod.
Root cause: Docking / rack connector or cable fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: Intermittent signal with cable movement.
Root cause: Worn patient cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Module input board | Major service part | $$$ | Difficult |
| Module / Tram / pod docking connector | Major service part | $$ | Moderate |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio