Fujifilm · Model FCR PROFECT ONE Plus
A multi-purpose FCR. Enhanced image processing. Superior image quality with 20 pixel/mm scanning pitch and Fujifilm's revolutionary. Dual-Side Reading Technology. High-precision images for various needs Dual-side reading technology The Dual-Side Imaging Plate (IP) Reading technology allows?the use of a thicker phosphor layer on the transparent base, ?thereby increasing DQE (Detective Quantum Efficiency) by collecting the emissions from both sides of the IP. High-precision Images for various needs. As High DQE is realized by the Fujifilm's Dual-side Reading Technology, with its low dose exposure, the FCR PROFECT CS Plus and the FCR PROFECT ONE Plus relieve patients' burden during X-ray examinations. Incorporated with the FUJIFILM proprietary image processing technology (Image Intelligence™), high resolution image reading with the 20 pixel/mm sampling pitch delivers excellent quality images in versatile diagnosis areas. For example, they are suitable for diagnosis areas requiring high-quality images even with low dose exposure, such as mammography, pediatrics, follow-ups and tests using a catheter. A multi-purpose FCR The FCR PROFECT ONE Plus now enables processing of pantomography exposures* (15 × 30 cm), in?addition to standard and long-view exposures, making a wider range of examinations possible. Enhanced image processing Image Intelligence™ a set of sophisticated digital image-processing software technologies available through Fujifilm's Console processes image data and optimizes final output.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Banding, streaks, ghost images, non-uniform density, dust artefacts.
Root cause: Dirty light guide / scanning optics, a worn erase-lamp array, imaging-plate wear, or contaminated transport rollers.
Symptoms: Plate jams in the reader, transport error, plate not returned.
Root cause: Worn or contaminated transport rollers / belts, or a sensor fault.
Symptoms: A faint previous image appears on the next exposure.
Root cause: Erase-lamp array degradation or a driver fault.
Symptoms: No or very noisy image, scan-error message.
Root cause: Scanning laser or the photomultiplier / detection chain.
Symptoms: Images will not display, annotate or send to PACS.
Root cause: Workstation PC or the DICOM configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transport rollers / belts | Periodic replacement | $$ | Moderate |
| Erase-lamp array | Periodic replacement | $$$ | Moderate |
| Imaging plates / cassettes | Periodic replacement | $$$ | Easy |
| Light guide / scanning optics | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.