Fresenius · Model PT, ST, VS
• Flow rate range 1 to 1000 ml/h (1500ml/h Optima ST) in normal mode. 0.1 to 100 ml/h in micro-infusion mode • Flow rate accuracy ± 5% with recommended sets • Volume range 1 to 9999 ml in normal mode. 0.1 to 999.9 ml in micro-infusion mode • Infusion time adjustable from 1 minute to 96 hours • Setting modes flow rate only, flow rate + volume, volume/time, rate + time, ramp up/ramp down, primary/secondary, sequential, bolus, induction/loading dose, micro-infusion • Optima VS: OCS (Occlusivity Check System) • Optima PT/ST: Drop sensor • DPS (Dynamic Pressure System): detection of pressure variations in the line allowing both a faster occlusion detection or a line disconnection detection • Anti-bolus system: automatic bolus reduction at occlusion release • Pressure limit adjustable from 100 to 900 mmHg • RS232 bi-directional connection • Battery life: 5 h 30 in average at 125 ml/h
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Uncontrolled flow when door open.
Root cause: Worn roller or failed free-flow valve.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable; per patient.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| IV administration set Consumable; per patient. | Consumable | $ | Easy |
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →