Fanem · Model Bilitron 3006
Microprocessor control of the radiance applied to the patient Alphanumeric display and easy-to-operate membrane keypad Emission in the blue light spectrum for treating bilirubin High radiance in the center and edges of the light focus No infrared or ultraviolet light emission Set of five SuperLEDs Average service life of 20,000 hours.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blue-light output weak or absent, irradiance below the therapeutic level on a radiometer, lamp-hours alarm.
Root cause: LED array / fluorescent tubes / halogen bulb at end of life, a driver / ballast fault, or a fibre-optic pad fault (blanket).
Symptoms: One section dim, patchy coverage.
Root cause: A failed LED segment or tube, or a damaged fibre-optic pad.
Symptoms: Overheating, thermal shutdown, loud fan.
Root cause: Failed or blocked cooling fan.
Symptoms: Treatment timer or lamp-hour meter wrong, controls unresponsive.
Root cause: Control-board fault.
Symptoms: Meter disagrees with a reference radiometer.
Root cause: Meter drift - needs calibration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| LED array / fluorescent tubes / bulb | Frequent wear item | $$ | Fairly easy |
| Driver / ballast | Periodic replacement | $$ | Moderate |
| Fibre-optic pad (blanket units) | Frequent wear item | $$$ | Fairly easy |
| Cooling fan | Periodic replacement | $ | Fairly easy |
| Control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.