NeoMedLight · Model BiliCocoon Phototherapy Light
For maximum effectiveness, it offers an intensive and homogeneous irradiance and a large coverage of body surface area It is emitting light in the absorption spectrum of bilirubin, facilitating the reduction of bilirubin concentration in the blood. A safe and controlled therapy, Cocooning including breastfeeding, Flexibility in usage in different settings both at hospital and at home, A potential savings through optimized workflow and lighter burden on staff.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blue-light output weak or absent, irradiance below the therapeutic level on a radiometer, lamp-hours alarm.
Root cause: LED array / fluorescent tubes / halogen bulb at end of life, a driver / ballast fault, or a fibre-optic pad fault (blanket).
Symptoms: One section dim, patchy coverage.
Root cause: A failed LED segment or tube, or a damaged fibre-optic pad.
Symptoms: Overheating, thermal shutdown, loud fan.
Root cause: Failed or blocked cooling fan.
Symptoms: Treatment timer or lamp-hour meter wrong, controls unresponsive.
Root cause: Control-board fault.
Symptoms: Meter disagrees with a reference radiometer.
Root cause: Meter drift - needs calibration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| LED array / fluorescent tubes / bulb | Frequent wear item | $$ | Fairly easy |
| Driver / ballast | Periodic replacement | $$ | Moderate |
| Fibre-optic pad (blanket units) | Frequent wear item | $$$ | Fairly easy |
| Cooling fan | Periodic replacement | $ | Fairly easy |
| Control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.