Edan · Model C3A
LED cold lighting system reproduces an original view of the tissue, 24 LED lights (illumination >1600lx) Advanced and fast auto-focusing system ensures a clear image Remote capture control and user-friendly keyboard layout for one-hand operation Electronic green filter enhances vascular image display without compromise of illumination Acetic acid and iodine reaction test timer to precisely detect low/high-grade CIN Real-time display of magnification, 1-28x Powerful Colposcopic findings gallery for comparison analysis New innovation – R-way™ Evaluation Program, that supports cervical disease diagnosis guidance for clinical users DICOM 3.0 supportive.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Light out, dim, flickering, uneven field, colour shift.
Root cause: Halogen bulb or LED module / driver at end of life, or a light-guide / power fault.
Symptoms: Fine focus slips or binds, magnification changer sticks, image not sharp.
Root cause: Worn focus drive or a magnification-changer detent fault.
Symptoms: Head drifts down, will not hold position, arm joints loose.
Root cause: Spring-arm counterbalance or friction brakes out of adjustment.
Symptoms: Filter will not engage or is stuck in place.
Root cause: Worn filter mechanism.
Symptoms: No or poor image, capture software will not connect.
Root cause: Camera, its cable, or the imaging PC.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Illumination bulb / LED module & driver | Frequent wear item | $$ | Fairly easy |
| Focus-drive / magnification-changer kit | Major service part | $$ | Difficult |
| Spring-arm counterbalance / friction-brake kit | Major service part | $$ | Moderate |
| Filter wheel / mechanism | Major service part | $ | Moderate |
| Camera / video module (video colposcopes) | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.